Lesson 2 of 3 · Resolution, sampling and calibration
Resolve and calibrate a spectrum
23 minute lesson · 3 checks along the way
Your goal for this lessonDistinguish spectral sampling from resolution and explain how calibration and stray light affect a weak signal.
In this lesson 4 sections · 3 checks
Read & understand
A spectrum is measured through an instrument
FLORIS combines two imaging spectrometers behind common fore-optics. Their different spectral capabilities support the weak-emission measurement and its surrounding context. A pushbroom instrument records a strip across the ground; spacecraft motion supplies successive strips. Matching channels therefore involves spatial and temporal registration as well as wavelength. A spectral value is useful only when its scene, timing and response are understood.
Coppo and colleagues’ 2017 design paper distinguishes a 0.1 nm sampling requirement from a 0.3 nm resolution requirement in selected oxygen bands. Sampling is the spacing between reported spectral positions. Resolution describes the width of the instrument’s response to a very narrow input, here expressed as full width at half maximum, or FWHM. These historical requirements are not evidence that every later flight measurement achieved the same performance. The current ESA instrument description explains the architecture; validation must establish actual behavior.
Check 1 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Broaden one invented absorption line
Imagine a reference spectrum with continuum 100 and a Gaussian-shaped dip to 20 at its center. Its intrinsic FWHM is 0.15 nm. Measure it through a normalized Gaussian instrument response with FWHM 0.30 nm. Normalized means that the response weights integrate to one. Each reported value averages neighboring wavelengths rather than selecting one perfectly narrow wavelength. This is an invented line and an idealized response, not a measured FLORIS calibration curve.
Convolving the two Gaussians gives a dip width √(0.15²+0.30²)≈0.3354 nm. Its depth is reduced by 0.15/0.3354≈0.4472, so the original depth of 80 becomes about 35.7771 and the measured center rises to 64.2229. The integrated area missing below the continuum remains the same in this ideal convolution. The feature has become broader and shallower. Both changes matter to a retrieval that relies on spectral contrast.
Check 2 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
More sample positions do not undo broadening
At 0.1 nm spacing there are three sampling intervals across a nominal 0.3 nm response width. The extra positions help describe the blurred shape, but they are not three independent 0.1 nm resolution elements. Interpolating to 0.01 nm spacing would add display positions without restoring the original 0.15 nm line. Recovering finer structure would require additional information and an explicit inverse whose uncertainty and assumptions are tested.
A useful calibration therefore measures more than where the samples are placed. The response width, shape, wavelength assignment and changes with temperature all affect interpretation. Shifting a response or assigning the wrong wavelength can make an observed feature disagree with its model. That mismatch may leak into the inferred fluorescence. A narrow sampling interval alone cannot establish that the spectrum is accurate, correctly registered or free of stray light.
Sampling a broadened spectral feature
Original normalized Gaussian convolution. The sample spacing and response width are different quantities. The plotted line and response are invented examples, not a measured FLORIS calibration function.
Open full-size diagram in a new tab: Sampling a broadened spectral featureRead & understand
Offsets and stray light need different evidence
Return to the first lesson’s Eₒ=100, Eᵢ=20, Lₒ=32 and Lᵢ=8. Add an uncorrected 0.2 radiance offset to both measurements. The reflectance estimate stays 0.30 because subtraction removes the common offset, but F rises from 2 to 2.2. That is a 10% fluorescence bias. More independent samples can reduce some random noise; they do not automatically remove a stable offset that mimics the emitted component.
ESA describes a sunlit diffuser for radiometric calibration, a dark reference and known spectral lines for wavelength checks. These references answer different questions. Light scattered from bright wavelengths into a dark spectral feature is scene-dependent stray light; a dark exposure alone cannot characterize every such contribution. Spatial stray light can also mix bright neighboring areas into a darker target. Independent calibration, scene screening and propagation of residual errors are needed before a weak inferred signal becomes a physiological claim.
Check 3 of 3
10 pointsUse the reading above. You can try again without losing points.
Put it together
Your lesson progress
Each check connects to a part of the story. Complete all 3 to earn your 25-point lesson bonus.
Explore further · Optional
Go to the source
Explore the evidence behind this lesson. These optional readings and videos open in a new tab.
- Read: Coppo et al.: FLORIS instrument designPages 1–4, especially Table 2, distinguish sampling and spectral resolution. Historical design requirements are not achieved flight accuracy.
- Read: ESA: FLEX instrument and calibrationShared fore-optics, two spectrometers and distinct radiometric, dark and wavelength references.