Lesson 2 of 3 · GOES nonlinear calibration and reference quality
Test the calibration reference
16 minute lesson · 3 checks along the way
Your goal for this lessonUse a known nonlinear response to calibrate a scene, and diagnose why a contaminated reference can make a good measurement look wrong.
In this lesson 6 sections · 3 checks
Read & understand
The reference path is part of the measurement
An infrared detector produces counts, not a ready-made cloud temperature. Calibration associates those counts with radiance in a defined spectral channel. The older GOES Imager used views of space and an internal blackbody as references. The references give the processing system known signals to compare with its detector output, but their optical paths and assigned radiances must also be trustworthy.
A 2003 NOAA description discusses a midnight calibration problem: radiation from warm structures could reflect from the internal blackbody toward the detectors. The reference view then contained an extra contribution. The proposed correction used instrument-temperature history to check responsivity. This is evidence about a particular older Imager calibration problem, not a statement that every GOES generation or every nighttime observation has the same fault.
Read & understand
Keep the nonlinear response explicit
Use an original simplified calibration equation R = q × C² + m × C + b. C is the detector count and R is radiance in an invented, consistently defined unit U. The coefficients carry units U/count², U/count and U. Assume independent characterization has established q = 0.0001 U/count² and that it remains fixed. This leaves m and b to determine from two distinct reference counts.
The historical description also uses a quadratic response with a laboratory-determined coefficient, but its complete processing includes additional corrections and filtering. Our numbers are fictional. We will omit those effects, keep the space reference fixed and work within the stated count range. We will not turn U into kelvin: converting channel radiance to a temperature would require the appropriate spectral response and a separate physical calculation.
Check 1 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Solve the response and reserve a target for checking it
Let the space reference have C = 100 and R = 0 U, and the warm reference C = 700 and R = 60 U. Subtract the two equations: 60 = 0.0001 × (700² − 100²) + m × (700 − 100). The quadratic contribution is 48 U, leaving 12 U across 600 counts. Therefore m = 0.02 U/count, and substitution into the space equation gives b = −3 U.
A scene at C = 400 now has R = 0.0001 × 400² + 0.02 × 400 − 3 = 21 U. Suppose an independently characterized check target really is 21 U and produces those 400 counts. It was not used to fit m or b, so it tests the prediction away from the two references. A fit with no residual at its own reference points is useful, but is not independent validation.
Check 2 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
A straight line can pass both references and still fail
Discard the independently known curvature by forcing q = 0. The line through the same reference endpoints is R = 0.1 × C − 10. It returns 0 U at count 100 and 60 U at count 700, apparently passing both calibration checks. Yet at count 400 it predicts 30 U, which is 9 U above the independent target. Exact agreement at the fitted endpoints has hidden an incorrect response shape.
If q had been unknown, two reference values would not identify all three coefficients. Another independent constraint would be needed. More arithmetic precision would not solve that shortage of information. Conversely, an unexpected target residual does not uniquely prove one cause: reference error, detector changes or an inaccurate response model can all matter. Diagnose them with evidence rather than automatically adjusting a coefficient until every displayed number looks plausible.
Read & understand
Contaminate the warm view while leaving the scene unchanged
Return to the known q and the original correct response. In a controlled counterfactual, keep the scene at count 400 and space at count 100. Extra radiation in the warm-reference path raises its count to 740, but the processor still assigns it 60 U. Repeating the fit gives m = [60 − 0.0001 × (740² − 100²)]/640 = 0.00975, then b = −1.975.
The unchanged scene is now reported as 16 + 3.9 − 1.975 = 17.925 U. Its apparent fall from 21 U comes entirely from the altered calibration in this example. Under the original correct response, count 740 corresponds to 66.56 U: 6.56 U more than the assigned blackbody signal. That added radiance belongs to the reference path here; it is not evidence that the cloud itself has cooled.
Fit the references; check the scene
Original count/radiance examples, not real detector coefficients or a midnight-correction implementation. Space is fixed at count 100 and 0 U. The original warm reference is count 700 at 60 U; the contaminated warm reference is count 740, still assigned 60 U. Radiance is not converted into temperature.
Open full-size diagram in a new tab: Fit the references; check the sceneCheck 3 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Use a check that does not inherit the same error
The contaminated fit still passes its own assigned warm and space references. Rechecking only those points would miss the fault again. An unchanged independent target exposes the resulting scene error. Other evidence can include reference temperatures, instrument state, quality flags and comparison with an independently calibrated instrument. Agreement with a second product is less informative if it inherits the same calibration or reference error.
The proposed midnight correction used screened historical responsivity and optics-temperature relationships, with conditions controlling when the correction could operate. The toy fit above does not reproduce that algorithm or establish its flight performance. Its lesson is narrower and useful: a reference is a measured part of the system, and fitting it can transfer its error into every later scene. Preserve the calibration context alongside the reported radiance.
Put it together
Your lesson progress
Each check connects to a part of the story. Complete all 3 to earn your 25-point lesson bonus.
Explore further · Optional
Go to the source
Explore the evidence behind this lesson. These optional readings and videos open in a new tab.
- Read: GOES I–M DataBook, Revision 1 (1996)Consult the older Imager’s calibration references, navigation and registration descriptions. Its generation-specific performance tables are not specifications for modern ABI imagery.
- Read: Weinreb and Han (2003): Implementation of Midnight Blackbody Calibration CorrectionExamine the documented reference-path contamination concern and proposed use of optics-temperature history. The full correction includes more conditions than the original polynomial exercise in this lesson.