Lesson 3 of 4 · Calibration transfer and bounded uncertainty
Transfer a calibration with care
18 minute lesson · 3 checks along the way
Your goal for this lessonTrace a stale dark reference into a false spatial contrast and propagate bounded reference uncertainty into a scene interval.
In this lesson 3 sections · 3 checks
Read & understand
Expose a false contrast on an equal scene
Now observe a uniform fictional scene S=100 with two detector positions. Position A has gain 1 and current dark offset 4; position B has gain 0.5 and current dark offset 8. Their raw readings are X_A=104 and X_B=58. Different raw numbers do not imply different scenes because the positions have different responses. Using the correct gain and current offset gives 100 at both positions.
Suppose an older calibration still assigns offset 4 to both positions, while the gains remain correct. Position A recovers (104−4)/1=100. Position B recovers (58−4)/0.5=108. The processing introduces an eight-unit contrast into a scene that was explicitly uniform. Calling that contrast a surface change would confuse a calibration mismatch with the quantity being observed.
An old offset invents a contrast
Original two-position model. The false contrast comes from a stale additive reference with known, unchanged gains; these are not measured LAPAN detector values.
Open full-size diagram in a new tab: An old offset invents a contrastRead diagram data as a table
| Position | Gain | Current offset | Raw reading | Correct recovery | Using old offset 4 |
|---|---|---|---|---|---|
| A | 1 | 4 | 104 | 100 | 100 |
| B | 0.5 | 8 | 58 | 100 | 108 |
Original two-position model. The false contrast comes from a stale additive reference with known, unchanged gains; these are not measured LAPAN detector values.
Check 1 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Trace the mismatch through the inverse
Write the current raw sample as X=gS+d_current, but process it with d_old. The estimated scene becomes S_hat=(X−d_old)/g=S+(d_current−d_old)/g. Thus the signed scene error is the dark-reference difference divided by the gain. At position B the raw offset difference is +4, and dividing by 0.5 gives +8 scene units. The sign matters: an offset that is too high would bias the estimate downward instead.
This formula holds because the gain is assumed unchanged and known. If gain also drifts, replacing only the dark reference cannot generally repair the full response. A dark exposure helps assess the additive component, while illuminated known levels are needed to assess gain. Do not apply the simple offset-error formula to a different failure model without re-deriving the measurement equation.
Check 2 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Carry a bounded reference into the answer
Suppose position B still returns X=58 with known gain 0.5, but its current dark offset is only known to lie from 6 to 10. The recovered scene is then bounded by (58−10)/0.5=96 and (58−6)/0.5=104. Larger assumed offset means smaller recovered signal, so evaluate both endpoints in the correct order. The answer is an interval, not an exactly known value of 100.
This interval is conditional on the supplied bounds and exact gain. It is not a confidence interval or a probability statement: we have supplied no distribution for the dark reference. Nor does it cover unmodeled stray light, saturation or gain drift. Report the assumed range alongside the result, then identify which additional measurement could narrow that range.
Check 3 of 3
10 pointsUse the reading above. You can try again without losing points.