Lesson 3 of 3 · Mass-to-charge calibration and peak width
Calibrate dust-ion flight times
19 minute lesson · 3 checks along the way
Your goal for this lessonCalibrate a square-root time-of-flight scale, propagate peak width and distinguish a mass-to-charge feature from a unique chemical or source identification.
In this lesson 4 sections · 3 checks
Read & understand
Follow the ions produced by one grain
SUDA is Europa Clipper’s impact-ionization time-of-flight mass spectrometer. A grain strikes a target, producing ions that electric fields guide toward a detector. Timing then constrains mass-to-charge ratios, while other signals provide information about the incoming grain. The instrument paper describes reflectron ion optics, laboratory calibration and fitted spectral lines. The mass of an ion represented by a peak is different from the mass of the original dust grain, which can produce many ions and fragments.
Use an original idealized spectrum with positive charge q = ze. Define the dimensionless ratio μ = m/(z u), where u is the atomic mass constant, and write t = t₀ + k√μ. All times are in microseconds; k includes the fixed acceleration and effective path geometry. Equal electric potential energy per charge makes heavier mass per charge travel more slowly. Assume correctly identified calibration lines, stable optics and a common time reference. These invented timings and widths are not SUDA flight or calibration measurements.
Read & understand
Solve the offset and stretch before assigning an unknown peak
Two reference lines have μ = 4 at t = 5 and μ = 16 at t = 9. Their square roots are 2 and 4, so k = (9 − 5)/(4 − 2) = 2 microseconds and t₀ = 5 − 2 × 2 = 1 microsecond. An unidentified peak centred at t = 11 gives μ = [(11 − 1)/2]² = 25. Treating time as directly proportional to mass would fail because the relation is linear in square root of μ instead.
The offset matters: setting t₀ to zero while keeping k = 2 would give (11/2)² = 30.25, not 25. Two lines determine two ideal parameters, but their chemical assignments must already be justified. Additional known lines can test the fitted scale rather than merely repeating the fit. SUDA’s calibration handbook describes identifying reference peaks and refining their positions; its instrument paper explains that path geometry and voltages affect the stretch. A classroom constant is not a universal calibration for every impact.
Check 1 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Carry a measured time width through the square mapping
Suppose the peak at 11 microseconds has a symmetric time-domain full width at half maximum, Δt = 0.1 microsecond. Local propagation through μ = [(t − t₀)/k]² gives Δμ ≈ 2μΔt/(t − t₀) = 0.5 and resolving power μ/Δμ ≈ 50. The time half-maximum edges at 10.95 and 11.05 map exactly to 24.750625 and 25.250625. Their distances from 25 are slightly asymmetric, even though the original time interval was symmetric.
The mapped edge span is exactly 0.5 for this symmetric interval, but it describes those time-domain half-height points. A mass-domain density obtained by rebinning also needs a Jacobian and can have different half-height locations. Use the local expression as a narrow-line approximation, not a universal peak-fitting algorithm. A second candidate at μ = 25.4 arrives only about 0.079683 microsecond later, less than the stated 0.1-microsecond width. Two numerical labels therefore do not establish two cleanly separated measured lines; line shape, relative strength and noise still matter.
Calibrate timing before assigning mass-to-charge
Two reference lines calibrate offset and stretch. Finite time width and charge-state ambiguity survive calibration; the plotted Gaussian is an original teaching model, not a SUDA observation.
Open full-size diagram in a new tab: Calibrate timing before assigning mass-to-chargeCheck 2 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Repeat the calibration and preserve the unresolved identity
In an alternate spectrum, μ = 4 arrives at 7 microseconds and μ = 16 at 13, giving k = 3 and t₀ = 1. A peak at 22 then has μ = 49. For Δt = 0.15, the local mass-to-charge width is 0.7 and resolving power is about 70; the exact mapped edge interval is 48.650625–49.350625. A candidate at μ = 49.5 differs in arrival time by about 0.106871 microsecond, again smaller than the measured time width.
The ratio μ = 25 is compatible with a 25-u singly charged ion or a 50-u doubly charged ion under this model; μ = 49 similarly permits 49 u at charge +e or 98 u at +2e. These are mathematical examples, not proposed Europa species. Fragments, isobars and contamination require further evidence even after charge is constrained. A spectrum also does not locate the grain’s surface origin without trajectory information. Calibration can strengthen a composition inference while leaving source attribution and any biological interpretation unresolved.
Check 3 of 3
10 pointsUse the reading above. You can try again without losing points.
Put it together
Your lesson progress
Each check connects to a part of the story. Complete all 3 to earn your 25-point lesson bonus.
Explore further · Optional
Go to the source
Explore the evidence behind this lesson. These optional readings and videos open in a new tab.
- Read: SUDA instrument paper: mass scale and resolutionKempf and colleagues, 2025: sections 4.1.1, 4.1.3 and 4.1.4 cover timing calibration, line widths and quantitative peak fitting. The lesson’s ideal spectrum is original.
- Read: SUDA PDS calibration handbookRevision A, 22 November 2024: section 4.1–4.2, printed pages 19–23, explains shift/stretch parameters and reference-peak calibration. The lesson uses a separate positive-time convention.
- Read: NASA: SUDA measurement chainDescribes incoming-grain trajectory measurements, impact ionization and mass-to-charge timing; trajectory evidence remains distinct from spectral identification.