Lesson 3 of 3 · Spectral leakage, inversion and representative calibration
Correct leakage with a measured response
25 minute lesson · 3 checks along the way
Your goal for this lessonSolve a calibrated two-channel mixing problem and recognize when measurement errors make its inverse unreliable.
In this lesson 4 sections · 3 checks
Read & understand
Calibration must measure where light actually goes
Reconstruction does not end the measurement problem. A real instrument can mix light from different wavelengths or scene locations. NIST’s hyperspectral-calibration work uses controlled spectral inputs to characterize these responses, then tests performance with more complex scenes. These are general methods for evaluating imagers; the paper does not report a Kent Ridge calibration campaign.
Build an original linear two-channel model after subtracting known dark offsets. Let x₁ and x₂ be the desired input signals in consistent arbitrary units. The measured outputs are m₁=x₁+0.1x₂ and m₂=0.2x₁+x₂. The coefficients represent a calibrated response, not conserved fractions of photons: detector gains and units are included. The model assumes stable coefficients, unsaturated detectors and linear response over the tested range.
Read & understand
Use independent inputs to measure the mixing
A known input (10,0) produces outputs (10,2), identifying the first input’s contributions to both measured channels. A second input (0,10) produces (1,10), identifying the other column of the response. Testing only an equal-input source would not independently determine all four coefficients. Repeat measurements across brightness and relevant conditions to check the assumptions rather than assuming that one pair of exposures establishes a universal correction.
For a new observation (m₁,m₂)=(10.2,4), solve the coupled equations together. Their determinant is 1−0.1×0.2=0.98. Thus x₁=(m₁−0.1m₂)/0.98=10 and x₂=(m₂−0.2m₁)/0.98=2. Substitution gives 10+0.1×2=10.2 and 0.2×10+2=4. The measured ratio 4/10.2 is about 0.392; the corrected input ratio is 2/10=0.2.
Check 1 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
An inverse can amplify small measurement errors
Subtracting leakage once using the measured channels as if they were the true inputs would give 9.8 and 1.96. Both are 2% too low because each measured channel already contains the other’s contribution. Dividing by 0.98 completes the simultaneous solution. The fact that the ratio happens to remain 0.2 in this particular example does not make the absolute estimates correct.
Consider a different response: m₁=x₁+0.99x₂ and m₂=0.99x₁+x₂. Subtraction gives m₁−m₂=0.01(x₁−x₂). True inputs (1,1) produce (1.99,1.99), but measurement perturbations of +0.01 and −0.01 give (2.00,1.98). Exact inversion now gives (2,0). Small output errors have produced large errors in the separate inputs. An invertible matrix is not necessarily a reliable measurement: report uncertainty and investigate conditioning, additional measurements or an explicitly justified constraint.
Check 2 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Test contrast as well as uniform illumination
A separate spatial example has y₁=0.9x₁+0.1x₂ and y₂=0.1x₁+0.9x₂, where the subscripts now identify neighboring scene locations. A uniform scene (10,10) remains (10,10); the blur is invisible in that test. A sharp contrast (10,0) becomes (9,1), exposing the unwanted transfer into the dark location. Passing a uniform-source test does not establish that edges or small dark features will be measured correctly.
NIST’s proposed scene projector combines spatial patterns with controlled spectra. For an original validation plan, reserve known mixed and high-contrast scenes that were not used to fit the response coefficients, then compare corrected outputs with their reference values and uncertainties. Include the expected brightness range and temporal changes. This checks the measurement chain’s intended use while keeping calibration, validation and an attractive false-colour display as separate results.
Calibrated inversion has an uncertainty cost
Original linear response models assume known offsets, stable coefficients and no saturation. Coefficients include detector gains and are not photon fractions. The ill-conditioned example uses a different response matrix.
Open full-size diagram in a new tab: Calibrated inversion has an uncertainty costCheck 3 of 3
10 pointsUse the reading above. You can try again without losing points.
Put it together
Your lesson progress
Each check connects to a part of the story. Complete all 3 to earn your 25-point lesson bonus.
Explore further · Optional
Go to the source
Explore the evidence behind this lesson. These optional readings and videos open in a new tab.
- Read: NIST: hyperspectral imager characterization and calibrationAll four pages and figures reviewed. Spectral/spatial characterization and representative validation support the method; the two-channel matrices and calculations are original, not Kent Ridge calibration data.
- Read: Segert and colleagues: Kent Ridge 1 preflight designConnects the course to the mission’s experimental hyperspectral payloads. Ground and balloon examples and preflight specifications do not establish calibrated orbital products.