Lesson 2 of 3 · Mars Science Laboratory
Infer precursors through pyrolysis controls
19 minute lesson · 3 checks along the way
Your goal for this lessonUse controlled precursor additions to estimate a reaction yield, then bound an original sample amount when another precursor can produce the same measured gas.
In this lesson 4 sections · 3 checks
Read & understand
The oven is part of the chemical experiment
Curiosity’s Sample Analysis at Mars suite, SAM, examines gases released from samples, including gases produced during heating. A molecule entering its chromatograph need not have existed unchanged inside the rock. Instrument-team laboratory work shows that heated oxychlorine salts can participate in oxidation or chlorination. A 2025 study of Cumberland material combined procedural controls, chromatographic and mass-spectral comparisons, and laboratory analog experiments to investigate long-chain alkanes and possible precursors. Detection and reconstruction of an original compound are related but different tasks.
Consider a fictional laboratory analog inspired by that interpretive problem, not a SAM flight procedure. Prepare equal, homogeneous aliquots with identical matrix, handling and heating. Let P be an original precursor and R the monitored product. Define yield η as nanomoles of R produced per nanomole of P; each P can contribute at most one R in this model. A known detector response g converts product amount into signal units. An independently matched procedural background B is additive and constant. These assumptions exclude unmodeled matrix-generated background.
Read & understand
Use added precursor to estimate conversion
For the first fixture, g = 4 signal units per nanomole of R and B = 8 units. Add a known precursor amount z of 0, 10 and 20 nmol to three separate aliquots. Their signals are 48, 68 and 88 units. If added and original P experience the same constant yield, the model is S(z) = B + gη(P + z). The first 10 nmol addition raises the signal by 20 units, so η = 20/(4 × 10) = 0.5. The second equal addition produces the same increment, supporting linearity over this tested range.
The unspiked excess is 48 − 8 = 40 signal units, corresponding to 40/4 = 10 nmol of product R. Under the sole-precursor model, that requires P = 10/0.5 = 20 nmol. Reporting 10 nmol as the original precursor amount would ignore incomplete conversion. Added molecules may be more accessible than native material trapped in a mineral, however. Unless equal recovery is justified, a spike slope measures the added precursor’s response and cannot automatically correct the native sample.
Check 1 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Bound a second route to the same product
Now permit another original precursor Q to produce R with known yield β = 0.25. The model becomes S(z) = B + g[η(P + z) + βQ]. Keep Q unchanged across aliquots and assume it does not alter η. The added-P slope still gives η = 0.5, but the unspiked product balance is 10 = 0.5P + 0.25Q. If independent evidence bounds Q between 0 and 8 nmol, its contribution ranges from 0 to 2 nmol of R. The compatible P interval is therefore 16–20 nmol.
That interval is a deterministic model bound, not a statistical confidence interval. More precise signal readings cannot remove the ambiguity between two unknown precursor contributions by themselves. A selective measurement of Q, a different reaction pathway or a chemically discriminating product could add the missing constraint. Likewise, suppose the 20 nmol addition instead produced 98 units while the first two readings stayed 48 and 68. The line predicts 88, so constant yield, constant background or aliquot equivalence would need investigation before extrapolating to the original amount.
Calibrate a reaction before inferring its precursor
Controlled additions estimate a conversion response under matched conditions. A second route to the same product leaves a bounded range of original precursor amounts, even with exact readings. The values and protocol are fictional.
Open full-size diagram in a new tab: Calibrate a reaction before inferring its precursorCheck 2 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Test the controls with an alternate matrix
For an alternate fixture, g = 5, B = 12, and the signals at additions 0, 10 and 20 nmol are 27, 42 and 57. The slope is 1.5 signal units per nmol, giving η = 0.3. The original product amount is (27 − 12)/5 = 3 nmol and the sole-precursor estimate is 10 nmol. If Q is bounded from 0 to 10 nmol with β = 0.1, then 3 = 0.3P + 0.1Q gives P between approximately 6.667 and 10 nmol.
Blank uncertainty also survives the calculation. With no Q but B anywhere from 10 to 14 units in this alternate fixture, P ranges from 8.667 to 11.333 nmol at the fixed measured slope of 1.5. A larger blank means less sample-derived material. These controls support a conditional chemical inference; they do not establish a biological origin. The real SAM study preserved uncertainty about precursor chemistry, even while identifying products through several complementary lines of evidence.
Check 3 of 3
10 pointsUse the reading above. You can try again without losing points.
Put it together
Your lesson progress
Each check connects to a part of the story. Complete all 3 to earn your 25-point lesson bonus.