Lesson 1 of 3 · Force calibration, lever arms and baseline drift
Calibrate a small force
23 minute lesson · 3 checks along the way
Your goal for this lessonConvert a measured displacement into force using the calibration geometry, a time-matched baseline and stated uncertainty.
In this lesson 4 sections · 3 checks
Using archive links & definitions
Follow a linked name to explore its archive card. “Back to your lesson” brings you to the same part of the reading.
Read & understand
A launch and a force measurement answer different questions
NuX-1 launched on Transporter-3 on January 13, 2022. NuSpace’s three-unit satellite carried an Internet-of-Things experiment using LoRaWAN and an Aliena miniature Hall thruster. Exolaunch supplied launch integration. The retained communications survey introduces remote sensor delivery; this course follows the propulsion experiment into the problem of measuring a very small force. The photograph shows the launch shared with MDASat-1a. NuX-1 is enclosed inside the fairing, so its hardware cannot be identified in the image.
Aliena’s 2022 concept paper describes early MUSIC-SI measurements affected by alignment, feed-line forces and temperature drift. The authors presented their thrust figures as estimates and called for a better measuring system. Later MUSIC-SI versions and the hollow-cathode MUSIC-HM are different configurations. Their results cannot become measurements of the NuX-1 flight article simply because the products share a family name.
Transporter-3 launches from Cape Canaveral on January 13, 2022, carrying NuX-1 and MDASat-1a among its rideshare passengers. The spacecraft are concealed inside the fairing; this complete photograph records their shared launch, not NuX-1 hardware or thruster performance.
Look closely
What part of NuX-1’s story can this photograph directly show?
Read & understand
Calibrate the assembled balance and its geometry
A balance converts force into motion. In a simple lever model, a force F acting perpendicular to an arm of length L produces torque F×L. Calibration therefore needs the location of the applied reference force as well as the sensor reading. Let S be sensor displacement per reference force, measured with the assembled installation. If the reference arm is Lc and the thruster arm is Lt, the inferred thrust is F=(Lc/Lt)×Δx/S.
Consider an original teaching balance: a 40-micronewton reference applied at Lc=0.12 metres produces an 8-micrometre displacement. Thus S=8/40=0.2 micrometres per micronewton. The thruster acts at Lt=0.24 metres. A net displacement Δx=12 micrometres gives F=(0.12/0.24)×12/0.2=30 micronewtons. Omitting the arm ratio gives 60, twice the appropriate force. The longer thruster arm needs less force to make the same torque. These are invented values, not a reconstruction of Aliena’s balance.
Check 1 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Estimate the zero at the measurement time
A raw sensor position includes its unloaded baseline. Suppose independent unloaded readings are 2 micrometres at t=0 seconds and 8 micrometres at t=12 seconds. For this exercise, assume the baseline changes linearly and that the calibration sensitivity stays fixed. At t=4 seconds, the interpolated baseline is 2+(8−2)×4/12=4 micrometres. A raw loaded position of 16 therefore gives Δx=16−4=12, recovering the 30-micronewton result.
Subtracting the first baseline instead gives Δx=14 and F=35 micronewtons. That five-micronewton difference is a systematic effect under this model; taking more identical loaded samples will not remove it. Linear interpolation is an assumption to test. A sudden temperature change, changing tubing force or residual gas flow can make the baseline curve differently. A reading immediately after electrical shutdown is not necessarily a true no-thrust reference if gas is still leaving the device.
Geometry and a time-matched zero determine inferred force
Original lever and baseline model. The illustrated force applications are separate measurements; all numerical values are fictional.
Open full-size diagram in a new tab: Geometry and a time-matched zero determine inferred forceCheck 2 of 3
10 pointsUse the reading above. You can try again without losing points.
Read & understand
Separate baseline uncertainty from changing sensitivity
If the interpolated baseline is uncertain by ±1 micrometre while every other quantity is fixed, Δx lies between 11 and 13. The corresponding force interval is 27.5 to 32.5 micronewtons. This is a bound for that one stated uncertainty, not a complete error budget. Uncertainty in the reference force, arm lengths, fitted sensitivity and alignment would also need evaluation.
Now suppose a later calibration gives S=0.24 instead of 0.2, with the same geometry. The same net displacement of 12 gives 25 micronewtons. Correcting the zero does not correct this sensitivity change. A trustworthy comparison checks both baseline and calibration under relevant conditions, then reports what remains uncertain. NuX-1’s history makes the distinction practical: successful integration and launch establish important milestones, while a defensible thrust value needs its own measurement chain.
Check 3 of 3
10 pointsUse the reading above. You can try again without losing points.
Put it together
Your lesson progress
Each check connects to a part of the story. Complete all 3 to earn your 25-point lesson bonus.
Explore further · Optional
Go to the source
Explore the evidence behind this lesson. These optional readings and videos open in a new tab.
- Read: Exolaunch: NuX-1 integration and payloadsComplete prelaunch article reviewed; integration and intended payload scope are distinct from experimental outcomes.
- Read: Exolaunch: Transporter-3 missionCompleted mission overview and manifest identify NuX-1 and the January 13, 2022 launch.
- Read: Laterza and colleagues: MUSIC concept and testingComplete 14-page primary paper and Figures 1–17 reviewed. Section III distinguishes uncertain early measurements and a newer SI configuration from later HM work.
- Read: Polk and colleagues: thrust measurement practiceSelected complete Sections IV.A, V.A and VI–VIII reviewed, including calibration equations 20–26 and error analysis. Figures are not reproduced; the interpolation and numbers are original teaching assumptions.